| Step 1:
Label each strip on
its end tab to ensure identity should the strips become detached
from the plate frame during the assay.
Step 2:
Designate one well on the plate
and leave empty. This well will serve as a substrate blank.
Step 3:
Pipette 200 µl of standards #1-6
into duplicate wells.
Step 4:
Pipette 200 µl of each specimen
into duplicate wells.
Step 5:
Cover the microplate with a plate
sealer and incubate the plate for 30 minutes at 37° C.
Step 6:
Aspirate the contents of each
well and wash the wells 4 times with 1X Plate Wash Buffer. To wash,
fill the wells with 300 ul of 1X plate wash buffer and aspirate.
Perform 4 fill/aspirate cycles. After the final wash cycle,
thoroughly blot the plate by carefully striking the plate on a pad
of absorbent paper towels. Continue striking until no visible
droplets of Plate Wash Buffer are observed.
Step 7:
Pipette 100 µl of Detector
Antibody into each standard and specimen well. Do not add Detector
Antibody to the substrate blank well.
Step 8:
Cover the plate with a plate
sealer and incubate for 30 minutes at 37° C.
Step 9:
Wash the plate 4 times with Plate
Wash Buffer as described in Step 6.
Step 10:
Pipette 100 µl of Substrate into
each well including the substrate blank well.
Step 11:
Incubate the plate for 30 minutes
at room temperature. A blue color will develop in wells containing
mouse IgG.
Step 12:
Pipette 100 µl of Stop Solution
into each well. A color change from blue to yellow will occur.
Step 13:
Within 15 minutes, read the
optical density of each well at 450 nm using a microtiter
plate reader. |